Useful Function
56
Measurement Display Smoothing
Use when wishing to stabilize displayed values when the load current is periodically fluctuating at high speeds.
The displayed value will be the average value of data, including new data, measured within the past 2 seconds. This
is renewed approximately every 200ms.
(This method is known as the Moving Average Method)
Setting Procedures
Press The Function “FUNC.” key.
The FUNC. lamp will light.
The Setting Item Number will be displayed in the Number
Display (voltage) and the parameter will be displayed in
the Number Display (current).
Select Item number 11 with the VOLTAGE dial and
set Parameter to 1 with the CURRENT dial.
Press the FUNC. key to finish.
The "FUNC." lamp is turned off and the Number Display
returns to the measurement display mode.