Status Codes
Status Codes
Status codes are continuously displayed at the top
of the controller screen.
Table 4.1
Method Status Codes
Code
Meaning
Air Cool
The furnace air cool line
has been opened to cool
the furnace.
Calib
The TGA Calibration
program is running.
Closing
The furnace assembly is
closing.
Cold
The instrument heater
cannot supply heat fast
enough to keep up with
the thermal program. This
may be caused by a large
ballistic jump in the
program, a faulty heater,
or a faulty control thermo-
couple signal.
Complete
The thermal method has
finished.
(table continued)