TA I
NSTRUMENTS
TGA 2050
Maintenance & Diagnostics
5–4
Routine Maintenance
Inspection
Examine the instrument periodically to keep it
free of dust, debris, and moisture. Keep the
furnace area clean. Any sample spillage of
residue should be removed before the next
experiment.
Cleaning the
Instrument Keypad
You can clean the TGA instrument keypad as
often as you like. The keypad is covered with a
silk-screen Mylar* overlay that is reasonably
water resistant but not waterproof or resistant to
strong solvents or abrasives.
A household liquid glass cleaner and paper
towel are best for cleaning the instrument
keypad. Wet the towel, not the keypad, with the
glass cleaner, and then wipe off the keypad.
Cleaning the
Quartz Furnace Tube
If the TGA is used to evaluate materials using
an oxygen purge, the furnace should be
cleaned routinely to prevent build-up of volatile
hydrocarbon residues that could combust.
Do not touch the furnace sample tube with
your bare fingers. Skin oils may cause
devitrification of the quartz glass, resulting
in severely reduced sample tube life.
!
W
ARNING
!
W
ARNING
Содержание TGA 2050
Страница 14: ...Introducing the TGA 2050 1 2 TA INSTRUMENTS TGA 2050 ...
Страница 26: ...Introducing the TGA 2050 1 14 TA INSTRUMENTS TGA 2050 ...
Страница 28: ...Installing the TGA 2050 TA INSTRUMENTS TGA 2050 2 2 Starting the 2050 2 32 Shutting Down the Instrument 2 33 ...
Страница 51: ...TA INSTRUMENTS TGA 2050 2 25 Figure 2 15 Installing the Tare Hang Down Wire Installing the Hang Down Wires ...
Страница 60: ...Installing the TGA 2050 2 34 TA INSTRUMENTS TGA 2050 ...
Страница 73: ...TA INSTRUMENTS TGA 2050 3 13 Figure 3 5 Furnace in Fully Closed Position Running an Experiment ...
Страница 86: ...Technical Reference TA INSTRUMENTS TGA 2050 4 2 ...
Страница 104: ...Technical Reference TA INSTRUMENTS TGA 2050 4 20 ...
Страница 106: ...TA INSTRUMENTS TGA 2050 Maintenance Diagnostics 5 2 ...
Страница 130: ...Index TA INSTRUMENTS TGA 2050 I 6 ...