4–17
Table 4.1
(continued)
Code
Meaning
Initial
The temperature is being
equilibrated to the desired
set point. When the
temperature has reached
equilibrium, the status
will change to “Ready.”
Iso
The thermal program is
holding the current
temperature isothermally.
Jumping
The heater is jumping
ballistically to the set
point temperature.
Load
The TGA is loading a
sample pan onto the
balance.
No Power
No power is being applied
to the heater. Check the
heater fuse.
Opening
The furnace assembly is
opening.
Ready
The system has equili-
brated at the initial
temperature and is
ready to begin the next
segment. Use the Start
function to continue the
method.
(table continued)
Status Codes
Содержание TGA 2050
Страница 14: ...Introducing the TGA 2050 1 2 TA INSTRUMENTS TGA 2050 ...
Страница 26: ...Introducing the TGA 2050 1 14 TA INSTRUMENTS TGA 2050 ...
Страница 28: ...Installing the TGA 2050 TA INSTRUMENTS TGA 2050 2 2 Starting the 2050 2 32 Shutting Down the Instrument 2 33 ...
Страница 51: ...TA INSTRUMENTS TGA 2050 2 25 Figure 2 15 Installing the Tare Hang Down Wire Installing the Hang Down Wires ...
Страница 60: ...Installing the TGA 2050 2 34 TA INSTRUMENTS TGA 2050 ...
Страница 73: ...TA INSTRUMENTS TGA 2050 3 13 Figure 3 5 Furnace in Fully Closed Position Running an Experiment ...
Страница 86: ...Technical Reference TA INSTRUMENTS TGA 2050 4 2 ...
Страница 104: ...Technical Reference TA INSTRUMENTS TGA 2050 4 20 ...
Страница 106: ...TA INSTRUMENTS TGA 2050 Maintenance Diagnostics 5 2 ...
Страница 130: ...Index TA INSTRUMENTS TGA 2050 I 6 ...