Use in an
Oxygen-Free
Atmosphere
A few extra precautions are necessary to ensure
an oxygen-free environment.
Purge Gas System
Use a high purity, inert gas of grade 5 or
better. It may be necessary to use an
oxygen trap in-line, depending on the purity
grade.
Choose a 2-stage regulator of diaphragm
construction for high purity applications.
Use copper or stainless steel tubing from
the gas regulator, to the flowmeters, and to
the TGA purge inlet ports.
Allow the TGA to prepurge (under closed
conditions) for at least 30 minutes after first
turning on your purge gas. Increase the
standard purge rate during this time.
100 cc/minute flow into the balance
chamber and 100 cc/minute into the
furnace.
Use in an Oxygen-Free Atmosphere