EN-15
DRAFT
1� Press
to remove air from the
inflow tubing� The
icon indicates that
the pump is operating�
Note:
This step must be performed each time
a new Inflow Cassette is inserted into the
pump�
2� Once the air has been removed and tubing is
completely filled with fluid, close the pinch
clamps or stopcock�
3� Open the pinch clamps or stopcock to
proceed with pump operation�
4� Press
to start/stop the pump�
During a Procedure
Changing the Pressure and Flow
Note:
Follow these instructions to select or change pressure and flow settings if the user preference file does
not specify these settings� Adjusting the settings will override the default and user preference file settings�
The system will select the following default settings for the following application areas:
Joint
Default Settings
Knee
45 mmHg
Shoulder
50 mmHg
Hip
50 mmHg
Small
35 mmHg
Note:
Default pressure settings are only suggestions, and each procedure and each patient may require
different parameters� Select the optimal pressure based on the patient profile, including, but not limited to,
blood pressure, height, weight, age, and tissue quality�
1� If necessary, press and to increase
or decrease the Set Pressure (adjust in
increments of 5 mmHg from 15–150 mmHg)
and Flow or Suction (adjust in increments of
10% from 0–100%)�
In this scenario, the pump is operating in
Inflow/Outflow mode� The Set Pressure is set
at 45 mmHg, and the Suction is set at 50%�
Note:
A green box will surround the suction/
flow value when the pump reaches the flow
limit� The pump is operating in constant flow
mode� Once the flow limit is reached, it will
operate in constant flow mode� (The pump
will attempt to operate at the set pressure
until the flow limit is reached�)