Error Message
Cause
Possible Solution
FAULT
Instrument recognition error: The stopcock
on the cannula is closed, the lower clamp
on inflow tubeset is closed, or no hardware
is attached to the tubing�
• Check the inflow tubing
from the bags to the joint
and ensure the hardware is
properly attached�
• Open any closed clamps or
stopcocks�
• Repeat the priming step�
FAULT
Inflow and/or Outflow Cassette not fully
inserted�
• Remove and reinsert the
cassette until it clicks into
place�
FAULT
Cassette is expired (after 24 hours of use)�
• Replace the cassette�
FAULT
The connection to the Crossfire console or
approved resection console is lost�
• Ensure the consoles are
properly connected to the
pump� Follow the instructions
in the “Setup and Device
Configuration” section in this
manual�
• Reboot the system�
• If the problem persists, contact
your Stryker representative or
return the console for repair�
FAULT
The Crossfire console is specified in the
user preference file, but it is improperly or
not connected to the pump�
• Ensure the console is properly
connected to the pump� Follow
the instructions in the “Setup
and Device Configuration”
section in this manual�
• Reboot the system�
• If the problem persists, contact
your Stryker representative or
return the console for repair�
LOCKDOWN
Hardware fault detected
• Reboot the system�
• If the problem persists, contact
your Stryker representative or
return the console for repair�
LOCKDOWN
Motor defective
• Reboot the system�
• If the problem persists, contact
your Stryker representative or
return the console for repair�
LOCKDOWN
Pressure Transducer out of calibration:
If the pressure is at a value greater than
20 mmHg for five seconds when no
cassette is inserted�
• Reboot the system�
• If problem persists, contact
your Stryker representative�
LOCKDOWN
Possible Overpressure: The pressure is
sustained at 750 mmHg for more than one
second, or a hardware failure or pressure
transducer malfunction has occurred�
• Reboot the system�
• If the problem persists, contact
your Stryker representative or
return the console for repair�