EN-34
DRAFT
h
1 m / 39 in
• Gather the following equipment:
• disposable Inflow Cassette tube set
• container filled with water
• fluid bag (3 L)
• Follow the instructions in this manual to power on the pump,
insert the Inflow Cassette, select a preference file, and select a
joint�
• Suspend a fluid bag 1 m/39 in on its holder and connect the
bag to the tubing according to the diagram�
• Discard the protective cap on the tubing (if needed), and
immerse the end of the tubing into a container filled with
water�
• Set the Pressure to 50 mmHg and Flow to 20%, according to the
instructions in the “Set the Pressure and Flow” section of this
manual�
• Remove air from the inflow tubing according to the instructions
in the “Prime the Inflow Tubing” section of this manual�
• Press
to start the pump� Allow the tubing to
completely fill with fluid�
• Press
to stop the pump� The actual pressure display
will show approximately 0–5 mmHg�
• Remove the tubing from the container and hold it at a given
water column height (h) indicated in the table below� Ensure
the tubing in the water column is completely filled with fluid�
• Read the actual pressure displayed on the pump� The test has
been successfully completed if the actual pressure on the pump
is within the range indicated in the table below for a given
height of the water column�
Height of water column (h)
Acceptable actual
pressure (mmHg)
Inches
Centimeters
12
30
20–25
18
45
30–35
24
60
40–45
46
90
65–70
5�
Inflow.Flow.Rate
• Gather the following equipment:
• disposable Inflow Cassette tube set
• fluid bag (3 L)
• stopwatch
• one 1 L measuring cup
• Follow the instructions in this manual to power on the pump, insert the Inflow Cassette, select a
preference file, and select a joint�
• Suspend a fluid bag 1 m/39 in on its holder and connect the bag to the tubing according to the
diagram�
• Discard the protective cap on the tubing (if needed), and place the end of the tubing into a container
filled with water�
• Set the pressure to 150 mmHg and flow to 50%�
• Remove air from the inflow tubing according to the instructions in the “Prime the Inflow Tubing”
section of this manual�
• Clamp off the tubing in the measuring cup without stopping the pump�