EN-20
DRAFT
Menu Features
Opening and Closing the Main Menu
1� If the pump is running, press
to
stop the pump�
2� Press
to open the Main Menu�
3� Press
to close the Main Menu�
Programming the Wash and Clear Functions Settings
Note:
Adjusting the settings will override the default and user preference file settings�
1� Press
to open the Main Menu�
2� Press
to open the Wash and Clear Settings Menu�
Setting
Function
Wash
• Inflow-only mode: Increases set pressure and flow limit by user-specified percentage over
user-specified duration� By default, the set pressure will increase by 50% of the current
setting, and the flow limit will increase by 100% of the current setting for 30 seconds�
• Inflow/Outflow mode: Increases set pressure and suction by user-specified percentage
over user-specified duration� By default, the set pressure will increase by 50% of the current
setting, and the suction will increase by 100% of the current setting for 30 seconds�
Clear
• Inflow-only mode: Increases flow limit by user-specified percentage over user-specified
duration� By default, the flow limit will increase by 100% of the current setting for 30
seconds�
• Inflow/Outflow mode: Increases suction rate by user-specified percentage over user-
specified duration� By default, the outflow will increaseby 100% of the current setting for
30 seconds�