EN-18
DRAFT
1� Press
to change to option
� A green
box will briefly appear displaying the new
hardware combination selection�
2� Ensure the correct hardware combination
is displayed� If the incorrect hardware
combination selection is displayed, or to
change the hardware combination selection,
follow the instructions in the “Change
Hardware Combination Selections” section�
3� Press
to change to hardware
combination
�
Performing the Wash Function
The Wash function increases the set pressure and flow limit by a user-specified percentage over a user-
specified duration (for Inflow-only mode), or increases set pressure and suction by a user-specified
percentage over a user-specified duration (for Inflow/Outflow mode)�
1� While the pump is running, press
to
perform the Wash function� Follow the
instructions in the “Program the Wash and
Clear Function Settings” section to adjust the
default settings�
2� Press
to repeat or stop the Wash
function�
Performing the Drain Function
The Drain function is only available in the Inflow/Outflow mode� It operates the outflow pump to remove
fluid from the surgical site for 30 seconds or until the user stops the pump�
1� Press
to stop the pump�
2� Press
to remove fluid from the joint�
3� Press
to repeat or stop the Drain function�
After a Procedure
Remove the Cassettes
Caution
•
Do not remove the cassettes while the pump is in use� The pump or the cassette may be damaged�
The cassettes can only be removed when the pump is stopped�
•
Do not attempt to remove the Outflow Cassette if it gets stuck as it may damage the pump or the
cassette� Follow the instructions in the “Troubleshooting” section in this manual to resolve this
problem�
1� Press
to stop the pump�
2� Close all pinch clamps�