![Steris Cmax Скачать руководство пользователя страница 90](http://html.mh-extra.com/html/steris/cmax/cmax_operators-manual_1374336090.webp)
4-1
764333-268
4.1 GENERAL
In addition to the routine maintenance, regularly
scheduled preventive maintenance is essential for safe
and reliable operation of the Cmax
™
Surgical Table.
Procedures described in this section of the
Maintenance Manual should be performed at regular
intervals as indicated in
,
P
REVENTIVE
M
AINTENANCE
S
CHEDULE
. The interval frequency
should be increased with increased usage of the
equipment. Should a problem occur, refer to
T
ROUBLESHOOTING
G
UIDE
.
A preventive maintenance program is available from
STERIS to help ensure peak performance and to avoid
unscheduled downtime. The program includes
maintenance adjustments and replacement of worn
parts by a qualified technician on a scheduled basis.
Contact STERIS for details.
Important:
If table is to be placed in extended storage,
have table prepared for storage by a qualified service
technician. Ensure batteries are disconnected and
check batteries before reconnecting. If table remains
in extended storage for longer than three months, table
must be operated through all articulations and the
batteries charged every three months.
4.2 PREVENTIVE MAINTENANCE
SCHEDULE
as a Preventive Maintenance record and,
along with the instructions in this section, as a guide
to performing preventive maintenance. Preventive
maintenance is not covered under warranty.
NOTE: For preventive maintenance, note the following:
1)
Recommended frequency of inspection is
yearly. Usage/Utility conditions may necessitate
fewer or more frequent inspections.
2) If repair or replacement is necessary, refer to
, for instructions.
3) Never permit unqualified persons to service
this equipment.
4) Maintain a record of all maintenance
procedures performed on this table.
4.3 CLEANING TABLE
4.3.1 General
NOTE: The user must follow the requirements of the
national committee responsible for hygiene and
disinfection when cleaning table.
See
for recommended cleaning products.
Clean the entire table per the following procedures.
Use the following materials to perform the cleaning
procedures described in this section:
• Several clean, dry, lint-free cloths.
WARNING – PERSONAL INJURY AND/OR
EQUIPMENT DAMAGE HAZARD: Repairs
and adjustments to this equipment must be
made only by STERIS or STERIS-trained
service personnel. Maintenance performed
by unqualified personnel or installation of
unauthorized parts could cause personal
injury, result in improper equipment
performance, invalidate the warranty or
result in costly damage. Contact STERIS
regarding service options.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: Use of incorrect hydraulic oil
may severely damage the table and/or
cause malfunction. Contact STERIS for the
proper hydraulic oil.
WARNING – PERSONAL INJURY AND/OR
EQUIPMENT DAMAGE HAZARD: Regularly
scheduled preventive maintenance is
required for safe and reliable operation of
this equipment. Contact STERIS to
schedule preventive maintenance.
WARNING – INFECTION HAZARD:
• To protect against aerosols being
reflected from potentially contaminated
surfaces, wear rubber or plastic gloves,
masks and eye protection, and follow
OSHA blood-borne pathogens
standards when cleaning.
• When cleaning/disinfecting table, do not
use alcohol which does not have sufficient
cleaning/disinfection properties.
WARNING – PERSONAL INJURY HAZARD:
When cleaning/disinfecting table, do not
use phenolics which may cause patient
skin burns if inadequately rinsed off, or
alcohol, which does not have sufficient
cleaning/disinfecting properties.
Section 4: Inspection and Maintenance
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......