Operating Manual
Safety Precautions
WARNING – PERSONAL INJURY HAZARD:
WARNING – INSTABILITY HAZARD:
WARNING – PINCHING AND TIPPING HAZARD:
Failure to keep the patient properly secured with the patient safety straps at all times could result in
death or serious injury to the patient and the operating room staff.
The floor locks must be engaged at all times when a patient is on the table. Failure to engage the floor
locks before placing a patient on the table or disengaging the floor locks while a patient is on the table
could result in the table rolling unexpectedly during the procedure.
Keep hands and feet clear of the unloading platform and the table base when unloading the table.
Serious personal injury could result.
Section 4.1,
Steps 1-6
in "Operating Table" must be executed before patient transfer to the table.
Execution of
Steps 1-6
with the patient on the table could result in injury to both the patient and the
operating room staff.
Healthcare professionals must ensure patients are positioned and monitored so as to prevent
compromising respiration, nerve pathways or circulation.
When installing any table accessory, check for correct attachment and tighten securely (if appropriate).
Do not use a worn or damaged accessory. Check installation before using any accessory.
Unanticipated table movement could cause patient injury. Patient must be secured to the table in
accordance with recommended positioning practices.
When cleaning/disinfecting table, do not use phenolics which may cause patient skin burns if
inadequately rinsed off, or alcohol, which does not have sufficient cleaning/disinfecting properties.
The table leg section is designed to support the patient legs.
Do not put body weight on the leg
section
.
If the integrity of the external protective ground installation or arrangement is in doubt, operate the table
from its internal power source.
If an antistatic path is necessary, STERIS recommends antistatic pads (specifically developed for this
table) in direct contact with the patient. Table must also be positioned on antistatic floor or connected to
equalization device (equipotential connector).
Do not remove table leg section if patient is not properly positioned. Patient
'
s torso and buttocks must
be firmly held in place by table back and seat sections and legs supported by leg supports.
Possible patient or user injury, as well as table or accessory failure, may result from using STERIS table
accessories for other than their stated purpose – or from using, on STERIS tables, accessories
manufactured and sold by other companies.
Patient injury may result if the operator of this table is not completely familiar with the controls for patient
positioning and table operation.
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......