Power-OnError
As the Surgical Table powers-on, the following compatibility errors may
Messages
Beach Chair Mode For
5085 General Surgical
Table
occur:
•
TABLE NOT CONNECTED - The Universal Hand Control did not
detect a connection to a Surgical Table.
•
TABLE NOT SUPPORTED -The Universal Hand Control is
connected to a Surgical Table that is not supported. See
I
NDICATIONS
F
OR
U
SE
of this publication for more information.
•
5085-SRT NOT SUPPORTED - The Universal Hand Control is
connected to 5085 SRT Surgical Table. See
I
NDICATIONS
F
OR
U
SE
of this
publication for more information.
•
BEACH CHAIR MODE ERROR – This message appears when
Universal Hand Control cannot cancel Beach Chair Mode for 5085 General
Surgical Table. See
B
EACH
C
HAIR
M
ODE
S
ECTION
of this publication for more
information.
Control of the Beach Chair Mode can only be accomplished with the
Advanced Hand Control. The table CANNOT function in the Beach Chair
Mode (see
Table 4
) using the Universal Hand Control configured for the
5085 General Surgical Table. The Universal Hand Control does not
incorporate the menu-driven interface that is supplied on the Advanced
Hand Control.
The Universal Hand Control cancels the Beach Chair Mode when it detects
that the 5085 General Surgical Table is in Beach Chair Mode. The message
EXITING BEACH CHAIR
appears on the display when Beach Chair Mode
cancellation is in progress (see Figure 9). Once the Beach Chair Mode is
cancelled, the Universal Hand Control displays normal operating symbols
on the display (see Figure 1) and begins normal table operation.
Figure 9. Message Displayed During 5085 Beach Chair Mode
Cancellation
If the Universal Hand Control detects a Surgical Table error is preventing the
cancellation of the Beach Chair Mode, the Universal Hand Control displays
the
SEE OPERATOR’S MANUAL
and
BEACH CHAIR MODE ERROR
messages on the display. Using the Universal Hand Control configured for a
Cmax/4085 also causes this error. A symbol appears in the lower right
corner of the display instructing the user to refer to the operator’s manual
(see Figure 10).
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......