7-5
764333-268
7. Seal along both seams of upper plastic column
covers as shown in Figure 7-12.
Figure 7-12. Seal Along Both Seams of Upper
Plastic Column Covers
8. After applying RTV, remove any excess sealant
using an alcohol dampened rag.
9. Allow sufficient cure time as recommended by the
RTV manufacturer before returning table for use.
7.5.2 Two-Part (Stainless-Steel) Shrouds
NOTE: To aid in loosening RTV between stainless-
steel base covers for future procedures, spray Dry
Lubricant Spray NG 202 (P752870-091) along the front
base shroud edge that contacts the RTV.
1. Seal mating areas between front and rear base
covers as shown in Figure 7-13.
NOTE: Ensure RTV is applied to both left and right
sides of base shrouds before assembly.
Figure 7-13. Seal Mating Areas Between Front
and Rear Base Covers
2. Apply RTV to bottom of each V-shaped cut-out
around interface with column shrouds (see
Figure 7-14).
Figure 7-14. Apply RTV to V-Shaped Cut-Outs
3. Apply RTV over two base shroud holes where
support bracket attaches (see Figure 7-15).
Figure 7-15. Apply RTV to Shroud Holes
4. Seal area where power supply inlet is exposed
through base (see Figure 7-6).
5. Seal around entire perimeter between upper plastic
column covers and column ensuring to fill entire
gap as shown in Figure 7-11. Smooth out RTV as
best as possible.
6. Seal along both seams of upper plastic column
covers as shown in Figure 7-12.
7. After applying RTV, remove any excess sealant
using an alcohol dampened rag.
8. Allow sufficient cure time as recommended by the
RTV manufacturer before returning table for use.
7.6 BATTERY REPLACEMENT
1. Turn Main Power Switch to OFF and remove
battery fuse (refer to supplied Operator Manual for
more information).
2. Remove Table Column Cover (see
3. Remove and save four screws (see Figure 7-4) at foot
end of base cover.
4. Remove stainless-steel base cover and set aside in
safe location.
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......