Page
12
of
14
Pressing any key while the
BEACH CHAIR MODE ERROR
message (see Figure 10) is displayed
returns the Universal Hand Control to normal operation (see Figure 1) with the following exceptions:
NOTE: The table should be serviced by a STERIS technician to resolve an
issue preventing cancellation of the Beach Chair Mode.
1
The 5085 General Surgical Table continues to operate in Beach
Chair Mode.
2
The 5085 General Surgical Table is operating in reverse orientation
and cannot be placed into normal orientation mode.
3
The ALS functionality for the leg section is inoperative to
accommodate a beach chair or shoulder accessory. If the leg
section is attached it can collide with obstacles.
4
Functionality of the 5085 General Surgical Table is allowed under
this condition so that the user can articulate the table to facilitate
handling the patient if necessary.
5
Using Universal Hand Control configured to a Cmax/4085 table also
causes this error. If a Cmax/4085 configuration was selected for the
Universal Hand Control Compatibility Mode, please select the 5085
Compatibility Mode when prompted.
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......