7-13
764333-268
12. Disconnect electrical connections.
Figure 7-32. Disconnect Hydraulic Lines From
Main Valve Block
13. Remove two fastening screws and replace main
valve block.
14. Reassemble by reversing procedure.
NOTE: When reassembling the hydraulic main valve
block, note the following:
1) Ensure all screws are secured with Loctite 242.
2) Replace (do not reuse) hydraulic screws and sealing
washers (see Figure 7-32). Ensure bolt threads,
manifold threads and hose connectors are free of
rubber seal particles.
3) Torque new screws to 14 ± 1.5 N•m (124 ±
13 lbf•in).
4) Follow Cmax Surgical Table Hydraulics figure in
the IPB Section of this manual to ensure proper
reconnection.
5) After reassembly, ensure all table functions operate
properly. Exercise all articulations at least five
times to completely purge system of trapped air.
Adjust movement speed if necessary (refer to
S
6) Check hydraulic oil level with table completely
lowered and tabletop level. Level should be 2-9/16"
(65 mm). Add oil (P764322-636) as needed.
7.20 REMOVE TABLETOP SECTIONS
Remove the Tabletop Sections as follows:
NOTE: Minor adjustments to accessory receptacle positions
may be made by loosening tabletop screws, inserting the
accessory and then torque the screws to 6 N•m (53.1 lbf•in).
7.20.1 Seat Section
Remove Tabletop Seat Section as follows:
1. Remove and save six slotted pan head screws
fastening tabletop to frame (see Figure 7-33).
2. Remove and set aside in safe area to avoid damage.
3. Reassemble in reverse order.
Figure 7-33. Remove Tabletop Seat Section
7.20.2 Kidney Section
Remove Tabletop Kidney Section as follows:
1. Remove and save four slotted pan head screws
fastening tabletop to frame (see Figure 7-34).
2. Remove and set aside in safe area to avoid damage.
3. Reassemble in reverse order.
Figure 7-34. Remove Tabletop Kidney Section
7.20.3 Back Section
Remove Tabletop Back Section as follows:
1. Remove and save 12 slotted pan head screws
fastening tabletop to frame (see Figure 7-35).
2. Remove and set aside in safe area to avoid damage.
3. Reassemble in reverse order.
Figure 7-35. Remove Tabletop Back Section
Lower
Fastening
Screw
Two-Way Hydraulic
Screw and Sealing
Washers
Hydraulic
Screw
and
Sealing
Washers
Upper
Fastening
Screw
Screws
Screw
Screws
Screws
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......