1-2
764333-268
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
WARNING – INFECTION HAZARD:
Do not store items on table base.
Doing so may result in equipment damage or inadvertent tabletop
movement placing the patient and/or user at risk of injury
.
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service
personnel. Maintenance performed by unqualified personnel or installation of unauthorized parts could
cause personal injury, result in improper equipment performance, invalidate the warranty or result in
costly damage. Contact STERIS regarding service options.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this
equipment. Contact STERIS to schedule preventive maintenance.
Table is factory set to operate at a certain voltage. Use of any other power supply could result in serious
personal injury and/or table damage
.
Use of this table in the presence of flammable anesthetics could result in serious personal injury and/
or table damage
.
Failure to perform periodic inspections of the table could result in serious personal injury and/or
equipment damage
.
During backup control, Tabletop functions are available even if table floor locks are NOT engaged.
LOCK table before moving tabletop or transferring patient
.
To prevent column cover from falling, securely hold cover when removing fastening screws
.
To avoid an action that could result in death or serious injury to the patient and the operating room
staff and/or serious equipment damage, completely evaluate the feasibility of every Auto Limit Sensors
(ALS) System suggestion taking into account the nature of the surgery and the surgical environment.
Ensure tabletop is properly supported before performing maintenance requiring tabletop to be
maintained in raised position
.
Ensure tabletop is adequately supported. When trendelenburg hydraulic cylinder is removed, tabletop
is free to move
.
Ensure tabletop is adequately supported. When tilt hydraulic cylinder is removed, tabletop is free to
move
.
Ensure tabletop is adequately supported. When hydraulic lines are disconnected from Main Hydraulic
Valve Block, tabletop is free to move
.
Lubrication makes the bearing rail assembly difficult to handle. Do not drop the assembly
.
To protect against aerosols being reflected from potentially contaminated surfaces, wear rubber or plastic
gloves, masks and eye protection, and follow OSHA blood-borne pathogens standards when cleaning.
When cleaning/disinfecting table, do not use alcohol which does not have sufficient cleaning/
disinfection properties.
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......