6-2
764333-268
Figure 6-1. Table Sensor Calibration
8. Display shows
ADJUST MODE.
NOTE: During table sensor calibration, note the
following:
1) The
REV. ORIENT
LED must be OFF.
2) Do not press
LEVEL
button during procedure
or control exits Calibration Mode
without
storing the values.
3) Pressing
FLOOR LOCK
button during
procedure orders the control to skip displayed
articulation. This is NOT recommended during
a final calibration.
9. If
ADJUST MODE
does not display or
REV.
ORIENT
LED is ON, turn Hand Control
OFF
then
ON
and repeat
Step 7
until correct results appear.
10. Press
REV. ORIENT
button, display shows
LEVEL ADJ
.
11. Using a level, position tabletop (starting with seat
section) so all sections are level in both horizontal
directions.
12. Slide tabletop until it is approximately 3/16" past
foot end of tilt frame (see Figure 6-1). This is
known as “Zero Slide” position.
13. Press
REV. ORIENT
button to save setting.
14. With display showing
TILT L MAX
, tilt tabletop
to maximum left tilt position and press
REV.
ORIENT
to save setting.
15. With display showing
TILT R MAX
, tilt tabletop
to maximum right tilt position and press
REV.
ORIENT
to save setting.
16. Display shows
TREND MAX
. Use Tilt Left button
to return table to center. Do not use
LEVEL
button.
17. Put table into maximum Trendelenburg then
press
REV. ORIENT
button to save setting.
18. Display shows
R TREND MAX
. Put table into
maximum Reverse Trendelenburg then press
REV. ORIENT
button to save setting.
19. Display shows
LEG D MAX
. Use
TREND
button
to return table to center. Do not use
LEVEL
button.
20. Lower leg section to maximum position and press
REV. ORIENT
button to save setting.
21. Display shows
LEG U MAX
. Raise leg section to
maximum position and press
REV. ORIENT
button to save setting.
Tilt frame
Extend table
top 3/16”
beyond the
foot end of the
tilt frame.
Tabletop
Tabletop
Extend Tabletop
3/16" (4.8 mm)
Beyond Foot
End of Tilt
Frame
Tilt Frame
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......