5-2
764333-268
Table 5-1. Troubleshooting Guide
Situation
Cause
Service Instructions
1. No green light on Main Power
Switch (if equipped with backlight).
• Main Power Switch is
OFF
.
• Power cord not plugged
into facility power.
• Defective power cord.
• Defective power switch.
• Turn Main Power Switch to
ON
.
• Plug power cord into appropriate facility
power supply.
• Replace power cord.
• Replace power switch.
2. No response from Hand Control or
Hand Control turns on momentarily
and then shuts off (may repeat two-
three times until there is no response;
power cord not plugged into facility
power).
• Battery fuse F8 open.
• Insufficient battery
charge.
• Defective Hand Control.
• Defective Hand Control
cable.
• Defective connector
plate/harness.
• Replace battery fuse F8.
• Connect power cord to appropriate
facility power and turn Main Power
Switch to ON. Allow batteries to charge
12 hours to restore full charge.
• Check operation using Backup Hand
Control (replace Main Control/cable).
• Check operation with Backup Hand
Control (replace hand connector plate/
harness).
3. No response from Hand Control
(power cord plugged into facility
power).
• Defective Primary Hand
Control/cable.
• Defective connector
plate/harness.
• Check operation using Backup Hand
Control (replace Main Control/cable).
• Check operation with Backup Hand
Control (replace hand control connector
plate/harness).
4. Cannot read Hand Control display
(power cord plugged into facility
power).
• Master Control Assembly
is malfunctioning from
defective Hand Control.
• Replace Hand Control and Master
Control Assembly.
5. When Hand Control is ON,
INTELLIPOWER Display bar graph
is OFF and Main Power Switch (if
equipped with backlight and power
cord is plugged into facility power) is
ON (green); or, INTELLIPOWER
Display bar graph and Battery Status
Indicator are ON while ac Power Plug
is OFF.
• Battery charger fuse F1
open.
• Faulty batteries.
• Faulty power supply.
• Replace battery charger fuse F1.
• Replace batteries.
• Replace power supply.
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......