![Steris Cmax Скачать руководство пользователя страница 91](http://html.mh-extra.com/html/steris/cmax/cmax_operators-manual_1374336091.webp)
4-2
764333-268
Table 4-1. Preventive Maintenance Schedule for the Cmax Surgical Table
Service Required
Minimum
Frequency
1.0 PREPARATION FOR PREVENTIVE MAINTENANCE
Discuss equipment operation with department personnel.
1x per year
Remove pads. Examine pad covers and fastening strips (pads and table).
1x per year
Examine clamps and other Side Rail hardware
.
Tighten all visible screws on Side Rails.
1x per year
Perform overall visual inspection of table, Hand Control and power cord.
1x per year
Check X-ray tops for standoffs tightness and signs of cracking (if equipped).
1x per year
Verify cap shroud integrity.
1x per year
Unfasten and remove table shrouds and base cover.
Each Inspection
2.0 HYDRAULIC SYSTEM
Check hydraulic oil filter element.
1x per year
Check hydraulic oil level.
6x per year
Inspect floor directly beneath table and all hoses, fittings and components of hydraulic
system for evidence of oil leaks.
1x per year
Verify operation of Auxiliary Control System foot pedal.
1x per year
3.0 CASTERS AND FLOOR LOCKS
Clean and inspect casters.
1x per year
Lubricate casters.
1x per year
Check Floor-Lock mechanism for proper operation.
1x per year
Fully extend floor lock cylinders and inspect for hydraulic leak (see
Each Inspection
Inspect table stability with Floor Locks engaged and on casters.
1x per year
Check pin and foot assemblies. Replace as needed.
1x per year
Verify all foot pads are present.
Each Inspection
Ensure table rolls freely and is easy to handle.
1x per year
4.0 CONTROLS
Inspect Hand Control housing and touch pads for damage.
Each Inspection
Using Hand Control and/or optional Physician-Controlled Foot Control,
check table positioning.
1x per year
5.0 ELECTRICAL CHECKS
Ensure all circuit board connectors and cable plugs are tight.
1x per year
Inspect main ac switch and ac power connector for damage.
1x per year
Содержание Cmax
Страница 1: ......
Страница 4: ......
Страница 65: ......
Страница 142: ...Page 3 of 14 ...
Страница 144: ...Page 5 of 14 Blue POINTER Indicates button user needs to push when selecting compatibility mode ...
Страница 148: ...Page 9 of 14 ...
Страница 149: ...Page 10 of 14 ...
Страница 152: ...Page 13 of 14 Feature Matrix Table 4 shows the functionality available for each compatible Surgical Table ...
Страница 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Страница 162: ...Page 7 of 8 e Reconnect the energy chain to the mounting bracket f Reconnect Slide Sensor Cable to slide rail mount ...
Страница 164: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......