background image

8

9

09.07.2019

The product must be recycled.

Only use indoors
WARNING! or CAUTION! or REMEMBER! 

Applicable to the given situation.  

(general warning sign)

Parameter description

Parameter value

Product name

Multimeter

Model

SBS-DM-1000

Battery type

9V

Rated power [W]

3

Protection rating IP

IP20

Over Voltage Protection

CATIII 1000V

Measuring range of direct/

alternating current voltage

200mV~1000V/ 2V~700V

Measuring range of 

direct/alternating current 

amperage

2mA~20A/ 2mA~20A

Measuring range of electri-

cal resistance/capacitance

200Ω~20MΩ/ 

2nF~200µF

Measuring range of 

temperature/frequency

-20~1000~/~200kHz

Working temperature that 

guarantees the highest 

measurement accuracy 

[° C]

23±5

Working temperature / 

Relative working humidity 

[°C, °F/%RH]

0~40°C/≤75%

Storage temperature/ 

Relative storage humidity 

[°C, °F/%RH] 

-20~60°C/≤80 %

Dimensions [mm]

195x89x40

Weight [kg]

0,32

TECHNICAL DATE

The terms "device" or "product" are used in the warnings 

and instructions to refer to < Multimeter >.

Do not use in very humid environments or in the direct 

vicinity of water tanks. Prevent the device from getting 

wet. Risk of electric shock! Do not cover the ventilation 

openings!

2.1. ELECTRICAL SAFETY

a) 

appropriate range.

b) 

Exercise caution when measuring voltage above 50V.

c)  If possible, avoid connecting the device to live 

circuits.

d) 

Before taking measurements of: resistance, diode or 

continuity test, capacitance test or temperature test, 

make sure that the tested circuit is not live.

e) 

Before taking the measurements, make sure that the 

correct function and measuring range have been 

selected.

f) 

Take special care in the event of a break in the circuit 

when using the device in a system with an electric 

current transformer connected to the terminals.

g) 

Before every use, make sure that the test leads and 

probes are in good condition, e.g. the insulation is 

undamaged.

h) 

Before disassembling the battery cover, disconnect 

the test leads and probes from the device and set the 

rotary switch to the "OFF" position.

i) 

When measuring, make sure that you are holding 

the insulated parts of the test leads. 

j) 

Please observe general electrotechnical principles 

and guidelines when taking measurements. 

U S E R   M A N U A L

1. GENERAL DESCRIPTION

The user manual is designed to assist in the safe and 

trouble-free use of the device. The product is designed 

and manufactured in accordance with strict technical 

guidelines, using state-of-the-art technologies and 

components. Additionally, it is produced in compliance 

with the most stringent quality standards

DO NOT USE THE DEVICE UNLESS YOU HAVE 

THOROUGHLY READ AND UNDERSTOOD THIS 

USER MANUAL.

To increase the product life of the device and to ensure 

trouble-free operation, use it in accordance with this user 

manual and regularly perform maintenance tasks. The 

technical  data  and  specifications  in  this  user  manual  are 

up to date. The manufacturer reserves the right to make 

changes associated with quality improvement. 

LEGEND

2. USAGE SAFETY

ATTENTION! 

Read all safety warnings and all 

instructions. Failure to follow the warnings and 

instructions may result in electric shock, fire and/or 

serious injury or even death.

The original operation manual is written in German. Other 

language versions are translations from the German.

PLEASE NOTE! 

Drawings in this manual are for 

illustration purposes only and in some details may 

differ from the actual product.

EN

2.2. SAFETY IN THE WORKPLACE

a) 

Make sure the workplace is clean and well lit. A messy 

or poorly lit workplace may lead to accidents. Try 

to think ahead, observe what is going on and use 

common sense when working with the device. 

b)  Do not use the device in a potentially explosive 

environment, for example in the presence of 

flammable  liquids,  gases  or  dust.  The  device 

generates sparks which may ignite dust or fumes.

c)  If you discover damage or irregular operation, 

immediately  switch  the  device  off  and  report  it  to 

a supervisor without delay.

d) 

If there are any doubts as to the correct operation 

of the device, contact the manufacturer's support 

service.

e) 

Only the manufacturer's service point may repair the 

The product satisfies the relevant safety 

standards.

Read instructions before use.

ATTENTION! 

Despite the safe design of the device 

and its protective features, and despite the use of 

additional elements protecting the operator, there is 

still a slight risk of accident or injury when using the 

device. Stay alert and use common sense when using 

the device.

3. DEVICE DESCRIPTION

The device measures the electrical values and the 

temperatures of selected devices and electrical circuits. 

The 

user is liable for any damage resulting from unintended 

use of the device.

3.1. DEVICE DESCRIPTION

3.2. PREPARING FOR USE 

The temperature of environment must not be higher than 

40°C and the relative humidity should be less  than 80%. 

Ensure good ventilation in the room in which the device 

is being used.

Note! The overload protection limits must not be exceeded.

Detailed electrical measurement parameters: 

• 

DC voltage

Overload protection: 1000V DC / 700V rms AC

Impedance: 10MΩ over the entire range.

ATTENTION! 

This product’s exploded view can be 

found on the last pages of the operating instructions 

(pp. 37). 

EN

Accuracy formula: ±% of r number of last 

digits  (at  temperature:  23  (±  5)  °C  and  humidity  ≤ 

75% RH).

1. 

LCD display 

2. 

On/Off button

3. 

HOLD button/Backlight: Lock/unlock the displayed 

value

4. 

Switching the display backlight and LED of a currently 

selected function on/off.

5. 

Diodes that inform about a presently active function.

6. 

Input jack: hFE

7. 

Knob: selecting the function and measuring range.

8. 

Mechanical locking of the jacks – protection against 

misuse.

9. 

Input jack:

a) 20A

b) mA

c)  VΩHz

, capacitance “+”, temperature “+”

d)  COM, capacitance “-”, temperature “-” 

device. Do not attempt any repairs independently!

f) 

In  case  of  fire,  use  a  powder  or  carbon  dioxide 

(CO2) fire extinguisher (one intended for use on live 

electrical devices) to put it out.

g) 

Children or unauthorised persons are forbidden to 

enter a work station. (A distraction may result in loss 

of control over the device)

h) 

Regularly inspect the condition of the safety labels. If 

the labels are illegible, they must be replaced. 

i)  Please keep this manual available for future 

reference. If this device is passed on to a third party, 

the manual must be passed on with it.

2.3. PERSONAL SAFETY

a) 

Do not use the device when tired, ill or under the 

influence of alcohol, narcotics or medication which 

can  significantly  impair  the  ability  to  operate  the 

device.

b) 

The device is not designed to be handled by persons 

(including children) with limited mental and sensory 

functions or persons lacking relevant experience 

and/or knowledge unless they are supervised by 

a person responsible for their safety or they have 

received instruction on how to operate the device.

c) 

When working with the device, use common sense 

and stay alert. Temporary loss of concentration while 

using the device may lead to serious injuries.

d) 

To prevent the device from accidentally switching on, 

make sure the switch is on the OFF position before 

connecting to a power source.

e) 

The device is not a toy. Children must be supervised 

to ensure that they do not play with the device.

2.4. SAFE DEVICE USE

a) 

Do not use the device if the ON/OFF switch does not 

function properly (does not switch the device on and 

off).  Devices  which  cannot  be  switched  on  and  off 

using the ON/OFF switch are hazardous, should not 

be operated and must be repaired.

b) 

When not in use, store in a safe place, away from 

children and people not familiar with the device who 

have not read the user manual. The device may pose 

a hazard in the hands of inexperienced users.

c) 

Keep the device in perfect technical condition. 

d) 

Keep the device out of the reach of children.

e) 

Device repair or maintenance should be carried out 

by qualified persons, only using original spare parts. 

This will ensure safe use.

f) 

To ensure the operational integrity of the device, do 

not remove factory-fitted guards and do not loosen 

any screws.

g)  Do not move, adjust or rotate the device in the 

course of work.

h) 

Do not leave this appliance unattended while it is in 

use.

i) 

Clean the device regularly to prevent stubborn grime 

from accumulating.

j) 

The device is not a toy. Cleaning and maintenance 

may not be carried out by children without 

supervision by an adult person. 

k) 

Never measure the voltage when the test leads are 

connected to the amperage measurement sockets.

m)  Exercise caution when working with voltages above 

DC60V or AC42V.

n) 

During measurement, hold the test leads or probes 

by the insulated parts.

n)  Select the appropriate function and measurement 

to avoid damage to the multimeter during the 

measurement.

o) 

Disconnect the test leads from the test points before 

switching to another function.

Range 

Accuracy

Resolution

200mV

±(0,5% +8)

100µV

2V

1mV

20V

10mV

200V

100mV

1000V

± (1,5% +8)

1V

Содержание SBS-DM-1000

Страница 1: ...NGSANLEITUNG MULTIMETER USER MANUAL MULTIMETER INSTRUKCJA OBSŁUGI MULTIMETR NÁVOD K POUŽITÍ MULTIMETR MANUEL D UTILISATION MULTIMÈTRE ISTRUZIONI PER L USO MULTIMETRO MANUAL DE INSTRUCCIONES MULTÍMETRO DE EN PL CZ FR IT ES ...

Страница 2: ...kreis nicht unter Spannung steht e Stellen Sie sicher dass die richtige Funktion und der richtige Messbereich ausgewählt sind bevor Sie die Messungen durchführen f Bei Verwendung des Gerätes mit angeschlossenem Stromwandler ist bei einem Stromausfall besondere Vorsicht geboten g Vor jedem Gebrauch sollte sichergestellt werden dass sich die Leitungen und Messungssensoren in einem guten technischen ...

Страница 3: ...it 75 RH 1 LCD Display 2 On Off Taste 3 HOLD Taste zur Aufrechterhaltung des Ergebnisses 4 Ein und Ausschalten der Beleuchtung des Displays 2 2 SICHERHEIT AM ARBEITSPLATZ a Halten Sie Ihren Arbeitsplatz sauber und gut beleuchtet Unordnung oder schlechte Beleuchtung können zu Unfällen führen Seien Sie voraussichtig beobachten Sie was getan wird und bewahren Sie Ihren gesunden Menschenverstand bei d...

Страница 4: ... die Sensorsonde in dem zu messenden Temperaturfeld Achtung Verwenden Sie für Temperaturmessungen immer eine Spezialsonde 3 3 1 9 DIODENMESSUNGUNDDURCHGANGSPRÜFUNG 1 Stellen Sie den Regler 7 auf den Messbereich 2 Verbinden Sie das schwarze Messkabel mit der Buchse COM und das rote mit der Buchse VΩHz 3 Schließen Sie die schwarzen und roten Messleitungen an die Kathode und Anode der zu prüfenden Di...

Страница 5: ...there is still a slight risk of accident or injury when using the device Stay alert and use common sense when using the device 3 DEVICE DESCRIPTION The device measures the electrical values and the temperatures of selected devices and electrical circuits The user is liable for any damage resulting from unintended use of the device 3 1 DEVICE DESCRIPTION 3 2 PREPARING FOR USE The temperature of env...

Страница 6: ...et to OFF and that the test leads are disconnected from the terminals m Before installing or replacing the batteries make sure that the device is not connected to an external circuit and that it is turned off SAFE REMOVAL OF BATTERIES AND RECHARGEABLE BATTERIES 9V batteries are installed in the devices Inserting a battery 1 Open the battery cover on the back of the device with a screwdriver 2 Inse...

Страница 7: ...a pomiarów odległości powierzchni oraz kubatury Odpowiedzialność za wszelkie szkody powstałe w wyniku użytkowania niezgodnego z przeznaczeniem ponosi użytkownik 3 1 OPIS URZĄDZENIA 3 2 PRZYGOTOWANIE DO PRACY Temperatura otoczenia nie może przekraczać 40 C a wilgotność względna nie powinna przekraczać 80 Należy pamiętać o tym by zasilanie urządzenia energią odpowiadało danym podanym na tabliczce zn...

Страница 8: ...d pomiarowy do gniazda COM a czerwony do gniazda mA dla minimum lub 20A dla maksimum 3 Podłączyć przewody pomiarowe szeregowo do obciążenia w celu wykonania pomiaru Uwaga Maksymalny mierzony prąd wynosi 200mA dla gniazda mA lub 20 A dla gniazda 20A Pomiar większej wartości będzie przyczyną przepalenia się bezpiecznika 3 3 1 5 REZYSTANCJA 1 Ustawić pokrętło 7 w preferowanym zakresie Ω 2 Podłączyć c...

Страница 9: ... do živých obvodů pod napětím d Před měřením odporu testem diod testem kontinuity nebo měření teploty se ujistěte že testovaný obvod není pod napětím e Před zahájením měření se ujistěte že byla vybrána správná funkce a rozsah měření f Buďte obzvláště opatrní dojde li k přerušení obvodu když zařízení používáte v systému s proudovým transformátorem připojeným ke kleštím g Před každým použitím zkontr...

Страница 10: ...né za jejich bezpečnost nebo od ní obdržely pokyny jak zařízení obsluhovat c Při práci se zařízením buďte pozorní řiďte se zdravým rozumem Chvilka nepozornosti při práci může vést k vážnému úrazu d Zabraňte náhodnému spuštění Ujistěte se že je spínač ve vypnuté poloze před připojením zařízení k napájecímu zdroji e Zařízení není hračka Dohlížejte na děti aby si nehrály se zařízením 2 4 BEZPEČNÉ POU...

Страница 11: ...ipojte měřící sondy v sérii k měřenému zdroji nebo zátěži Pozor Maximální naměřený proud je 200 mA pro zdířku mA nebo 20 A pro zdířku 20 A Měření větší hodnoty způsobí vyhoření pojistky 3 Připojte červený i černý měřící kabel ke katodě i anodě testované diody odečtěte pokles napětí z displeje 4 Připojte měřící kabely k obvodu jehož spojitost má být testována Pokud je odpor menší než 80 Ω bude akus...

Страница 12: ...e à la dernière page du manuel d utilisation p 37 1 Écran LCD 2 Bouton On Off 3 Bouton HOLD Rétroéclairage Verrouille déverrouille la valeur affichée 4 Allumer éteindre le rétroéclairage de l écran et de la diode correspondant à la fonction actuellement sélectionnée 5 Diode informant de la fonction actuellement active 6 Borne d entrée hFE 7 Sélecteur sélectionner la fonction et la plage de mesure ...

Страница 13: ...e Description Conditions de test Affichage de la tension approximative de la diode Courant continu environ 1 5 mA Tension du courant continu environ 3 0V La sonnerie intégrée retentit lorsque la résistance est inférieure niż 80 Ω Tension de circuit ouvert environ3 0V Plage Précision Résolution 2mA 0 8 8 1µA 20mA 0 8 8 10µA 200mA 1 2 8 100µA 20A 2 0 10 10mA Plage Précision Résolution 2mA 1 8 8 1µA ...

Страница 14: ...nF 200µF Campo di misurazione della temperatura frequenza 20 1000 200kHz Temperatura di esercizio per la massima precisione di misura C 23 5 Temperatura di esercizio Umidità relativa C F RH 0 40 C 75 Temperatura ambiente circostante Umidità relativa ambiente circostante C F RH 20 60 C 80 Dimensioni mm 195x89x40 Peso kg 0 32 DATI TECNICI Il termine apparecchio o prodotto nelle avvertenze e descrizi...

Страница 15: ... I bambini e le persone non autorizzate non devono essere presenti sul posto di lavoro La disattenzione può causare la perdita del controllo sul dispositivo h Controllare regolarmente lo stato delle etichette informative di sicurezza Se le etichette non sono ben leggibili devono essere sostituite i Conservare le istruzioni d uso per uso futuro Nel caso in cui il dispositivo venisse affidato a terz...

Страница 16: ...manopola 7 sul valore 2 Collegare il cavo di misura nero alla presa COM e quello rosso alla presa VΩHz 3 Collegare i cavi di misura neri e rossi al catodo e all anodo del diodo testato leggere sul display il valore della caduta di tensione 4 Collegare i puntali di misura al circuito di cui si vuole testare la continuità Se la resistenza sarà inferiore a 80Ω circa il segnale acustico sarà continuo ...

Страница 17: ...DESCRIPCIÓN DEL DISPOSITIVO ATTENTION Le plan de ce produit se trouve à la dernière page du manuel d utilisation p 37 1 Pantalla LCD 2 Botón On Off 3 Botón D HOLD luz de fondo Bloqueo desbloqueo del valor mostrado 4 Encender apagar la luz de fondo de la pantalla y los diodos de la función seleccionada actualmente 5 Diodos informando sobre la función actualmente activa 6 Ranura de entrada hFE 7 Man...

Страница 18: ... la toma VΩHz 3 Conecte los cables de prueba negro y rojo al cátodo y al ánodo del diodo que se va a probar lea la caída de voltaje en la pantalla 4 Conecte los cables de prueba al circuito cuya continuidad se va a probar La señal acústica será continua si la resistencia es inferior a aproximadamente 80Ω 3 3 1 10 PRUEBA DE TRANSISTOR HFE 1 Ajuste el mando 7 en el rango de medición hFe 2 Determine ...

Страница 19: ...e esté apagado ELIMINACIÓN SEGURA DE ACUMULADORES Y BATERÍAS En el aparato se utilizan baterías 9V Para montar la batería 1 Abra la tapa de la batería en la parte posterior del dispositivo con un destornillador 2 Inserte las pilas correspondientes a los parámetros técnicos 3 Cierre y atornille la tapa de la batería Retire las baterías usadas de la unidad siguiendo el mismo procedimiento que para l...

Страница 20: ...rter Nazwa produktu Model CZ Dovozce Název výrobku Model FR Importateur Nom du produit Modèle IT Importatore Nome del prodotto Modello ES Importador Nombre del producto Modelo 4 5 6 DE Batterietyp Produktionsjahr Ordnungsnummer EN Battery type Production year Serial No PL Typ baterii Rok produkcji Numer serii CZ Typ baterie Rok výroby Sériové číslo FR Type de batterie Année de production Numéro de...

Страница 21: ...Verfügung stehenden Möglichkeiten der Rückgabe oder Sammlung von Altgeräten 1 RICHTLINIE 2002 96 EG DES EUROPÄISCHEN PARLAMENTS UND DES RATES ÜBER ELEKTRO UND ELEKTRONIK ALTGERÄTE 2 Gesetz über das Inverkehrbringen die Rücknahme und die umweltverträgliche Entsorgung von Elektro und Elektronikgeräten Elektro und Elektronikgerätegesetz ElektroG Umwelt und Entsorgungshinweise Utylizacja produktu Prod...

Отзывы: