FEATURES OF THE SYSTEM BUILD
In the section measurement (fig. 8) select 'Distance' in the
drop down field. Beside it, choose which axis should be
measured.
Underneath, check the box depending on whether you wish
to choose between adjustment or laser-head only. See sec-
tions 6.4.3 and 6.4.8 for more information.
In the section 'Parameters' you may want to change the
averaging value in the drop down field. In most cases this
should be set to 0.1s. Also, for correct compensation in tem-
perature, make sure to select the right material in the drop
down field for the termal expansion coefficient (in the case
of machine tools this is usually steel).
At the bottom of the menu you can see the status panel. The following table explains the
individual bars:
Fig. 9: Parameters
Bar
Status
Fix
No connection to interferometer
Restart interferometer. Select connect in configura-
tion manually
Connection OK
No intervention needed
Laser switched off
Click bar to start laser
Interferometer booting
Wait for interferometer to finish booting
Critical operation
Finish current task. Click bar to allow laser to reach
stable operation
Fig. 8: Measurement