6
External input/output functions MG-S series operation manual
-
64
-
6-5
Input command
6-5-1
Transmission procedure
1
Send an input command from an external device to the balance.
The table below shows the enable/disable of input commands in each measuring mode.
Commands
Measuring
mode
Zero-point adjustment,
Tare subtraction,
Date/Time output
Output control,
Comparator setting,
Interval time setting
External contact input
Weighing
x
x
x
Counting
x
x
x
Percentage
x
x
x
Multiplied by
Coefficient
x
x
x
Specific gravity
x
-
x
Statistics
x
-
x
Animal
x
-
x
Formulation
-
-
-
2
Upon successful completion of an input command, the balance will send either a normal
completion response or the result data requested by the command to the external device.
- If the operation has not resulted in successful completion, or if the command is invalid (an error), the
balance will transmit an error response.
- When the balance is in normal display mode, it usually sends a response to a command within one
second of receiving the command. For the tare range, a response is sent after the commands are
completely processed.
(1) After you have sent an input command, the balance return the response command
approximately in 1 second.
(2) Do not send another command to the balance until the external device receives a response
from the balance.
(3) If the balance receives a command when you are setting a function, when the balance is
under span calibration, or the balance is busy for other reasons, the command is ignored.
In the case that <17 WT STABLE> is <ON>, the balance waits the weighing stability after receiving
Tare-subtraction command/Zero-point adjustment command.
For verified balance, <17 WT STABLE> is fixed to <ON> and the balance always waits
the weighing stability after receiving such a command.
Содержание MG-S Series
Страница 2: ......
Страница 10: ......
Страница 67: ...MG S series operation manual 6 External input output functions 57 ...
Страница 68: ...6 External input output functions MG S series operation manual 58 ...
Страница 84: ...8 Controlling and adjustment functions MG S series operation manual 74 ...
Страница 103: ...MG S series operation manual Appendix 93 Appendix 2 Dimensional outline drawing MG S322 MG S1501 MG S8200 ...
Страница 109: ...MG S series operation manual Appendix 99 Appendix 7 Windshield assembly instructions ...
Страница 110: ...Appendix MG S series operation manual 100 ...
Страница 111: ...MG S series operation manual Appendix 101 ...
Страница 113: ......
Страница 114: ......