A-4
Technical Data
Measurement Ranges and Tolerances
Essential Performance
The Merlin™ PSA provides diagnostic information related to the implanted lead and provides pacing
pulses at a rate and amplitude set by the operator.
The Merlin PSA operates in conjunction with the Merlin™ PCS Model 3650; therefore, the essential
performance defined for the Merlin PCS governs and limits the available performance that is exhibited
by the Merlin PSA EX3100.
Measurement
Range
Tolerance
ECG
±5 mV
0 to -20% of the vertical display gain
EGM
±35 mV
0 to -10% of the vertical display gain
Sensed intervals
100-1991 ms
±10 ms
Measured heart rate
30 - 300
min§
±1
min§
Intrinsic signal amplitude
0.1 - 35.0 mV
±0.1 mV or ± 10%
1
1. Whichever is greater
Slew rate
0.1 - 8.0 V/s
±0.1 V/s or ± 10%
Pacing lead impedance
200 - 4000
¬
±10% for impedances in the range of
200 - 2499
¬
(at 5.0 V)
±15% for impedances in the range of
2500 - 4000
¬
(at 5.0 V)
AV Delay
25 - 450 ms
±5 ms
Base Rate
30 - 220
min§
(2000-273 ms)
±5 ms
Interventricular Delay
10 - 80 ms
±5 ms for nominal pulse width
Pulse Amplitude
0.1 - 10.0 V
±5% or 0.08 V, whichever is greatest for
pace intervals > 400 ms (with a pacing load of
500
¬
)
±20% or 0.4 V, whichever is greatest for pace
intervals of 40-400 ms (with a pacing load of
500
¬
)
Pulse Width
0.1 - 1.5 ms
±0.015 ms
Table A-4. Ranges and tolerances for Merlin PSA measurements