3-8
Rhythm Display
Rhythm Display Setup Instructions
1. Select the Adjust Display button to the right of the Rhythm Display.
The
window appears.
2. Locate position 1.
3. Select the Source you want to see in position 1 (
,
, or Off).
The programmer selects a default Configuration for the Source.
4. Select the Configuration button.
If you selected ECG or EGM for the Source, the
, select the Basic or Full button.
5. Choose the configuration.
6. Repeat these steps for the remaining waveforms.
7. Select the Sweep Speed button and choose a speed.
8. Select the ECG Filter button to enable the ECG filter
Electromagnetic interference is reduced when the ECG filter is enabled.
9. Select the AutoGain All button to refresh the AutoGain settings.
Adjust Display
The Adjust Display window changes the:
•
Source
for each waveform in the Rhythm Display window (
•
Configuration
of the waveform
• Sweep Speed
•
ECG Filter
to reduce electromagnetic interference
The following buttons are also available:
•
AutoGain All
. Sets all waveforms to AutoGain and recalculates the gain of waveforms currently
displayed in the Rhythm Display.
See also:
•
•
•
Rhythm Display Setup Instructions
Accessed From: Rhythm Display > Adjust Display button
ECG Configuration
The ECG Configuration window changes the ECG vector on the
For an example of a typical ECG setup, refer to the Merlin PCS Model 3650 User’s Manual.
To achieve the ECG vectors, select the following electrodes:
•
I
. L(+) to R(-)
•
II
. F(+) to L(-)
•
III
. F(+) to R(-)
•
aVR
. R(+) to [L(-) + F(-)]
•
aVL
. L(+) to [R(-) + F(-)]