Merlin™ PSA Help Manual
A-5
Electromagnetic Compatibility
The Merlin PSA requires special precaution with regard to electromagnetic compatibility (EMC) and
should be used in accordance with the EMC information provided in this section of the manual.
Do not use the Merlin PSA near other electronic equipment. In particular, avoid portable and mobile RF
communication equipment, such as mobile phones, as well as equipment emitting strong magnetic
fields. This may affect the operation of the Merlin PSA.
The Merlin PSA complies with the requirements of the international EMC standard IEC 60601-1-2
when used with the cables listed in Table A-8 on page A-11.
The Merlin PSA is intended for use in the electromagnetic environment specified in Table A-5 and
Table A-6.
The Merlin PSA is intended for use in an electromagnetic environment where radiated RF disturbances
are controlled. Maintain a reasonable distance between portable and mobile RF communications
equipment (transmitters) and the Merlin PSA.
Note
For single fault conditions, see Troubleshooting (page A-18).
Test
Compliance
Electromagnetic Environment – Guidance
RF emission
CISPR 11
Group 1
The Merlin PSA uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emission
CISPR 11
Class A
The Merlin PSA is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not Applicable
Voltage
fluctuations/
flicker
emissions IEC
61000-3-3
Not applicable
Table A-5. Guidance & manufacturer’s declaration – electromagnetic emissions