Merlin™ PSA Help Manual
1-1
1. Introduction
Contents:
•
•
•
•
•
•
System Description
The Merlin™ Pacing System Analyzer (PSA) Model EX3100 system provides you with the ability to
assess the electrical performance of a lead system during pulse generator implantation or invasive
lead-system troubleshooting. The Merlin PSA system uses the St. Jude Medical™ Merlin™ Patient Care
System (PCS).
In addition to the Merlin PCS, the Merlin PSA system consists of:
• a sterilizable patient cable
The Merlin PSA Patient Cable Model EX3150 connects the Merlin PSA with one to three implanted
leads.
• patient-cable adapters
The Merlin PSA Cable Adapter Model EX3170 and the Merlin PSA “M” Adapter Model EX3180
connect the Merlin PSA to commonly used PSA patient cables (see
). The PSA patient cables are then connected to the implanted leads.
For information on the Merlin PCS, refer to the Merlin PCS Model 3650 User’s Manual. For technical
data and other information on the Merlin PSA and accessories, refer to
.
Operator Profile
The intended operators of the Merlin™ PSA system include trained St. Jude Medical employees and
clinical personnel trained in pacing and electrophysiology, for example, nurses, technicians, or
physicians.
Note
Only persons properly trained to operate the Merlin PSA should do
so.
If the user connects both the Merlin PSA Model EX3100 and the PSA
Wand Model 3150 to the Merlin PCS, the programmer only enables
the Merlin PSA Model EX3100.
Refer to the Merlin PCS Model 3650 User’s Manual for Merlin PSA
installation instructions.