4-2
Merlin™ PSA Session Information
Figure 4-1. Merlin PSA and approved patient cables and cable adapters
3. Inspect the Merlin PSA, cable, or cable adapter for any visual damage.
4. Insert the connector on the cable or cable adapter into the green PSA port on the
Merlin PSA.
Ensure that you have a secure connection.
5. If not already connected, insert the ECG into its port on the Merlin PSA.
6. If you are using a cable adapter (Models EX3170 or EX3180), connect the cable adapter to
its respective patient cable.
7. Connect the patient cable connectors to the implanted lead you wish to test.
For appropriate measurements:
- Connect the lead to the appropriate channel
- Connect the (+) and (–) patient-cable to the appropriate connection on the lead. Refer to the
appropriate lead manual for information on the lead configuration.
The clips on the Model EX3150 patient cable can be positioned side by side, with the lead elec-
trodes visible through the clip.
CAUTION
Do not attempt to insert a patient cable or cable adapter into the
PSA port on the right side of the Merlin PCS programmer.
EX3180
EX3150
4051L
EX3170
Merlin PSA
4161