A-18
Technical Data
Troubleshooting
Loss of Communication
If communication between the PSA software and the Merlin PSA device is broken, for example, due to
a cable disconnection, burst stimulation is terminated and you will no longer be able to change any
parameter settings. In case of a communication loss, the Merlin PSA stores the current parameter
settings and restores them when communication is restored. The Merlin PSA displays a lost
communications alert message. Programming is not possible when a break is detected.
To remedy a communications loss, turn off the MerlinPCS, pull the Merlin PSA unit from the storage
compartment, and re-install the Merlin PSA cable into the PSA port on the Merlin PCS.
System Failure
Each time the Merlin PCS is turned on, the Merlin PSA conducts a self-test. If the test fails, the Merlin
PCS will not begin a PSA session and the blue LED on the unit will not illuminate. Contact
If the Merlin PCS fails during an ongoing PSA session, the unit will resort to Safe Pacing mode. See
for more information.
Low Batteries
Batteries are required for backup operation in case of a system failure. See
information on replacing the batteries.
EGM Noise
To reduce noise on the EGM display caused by interference with the mains power:
• Do not coil the patient cable or patient cable adapters.
• Keep the patient cable and patient cable adapters away from the power cord.