8-6
Mode Descriptions
Figure 8-8. AOO Mode timing diagram
ODO
In this mode, pacing is turned off while the device continues to detect and record sensed events. This
mode is useful primarily for temporary diagnostic evaluation and recording of intrinsic activity.
CAUTION
AOO mode is primarily intended for temporary use. Long-term use
may result in competitive pacing, which may induce potentially
dangerous tachyarrhythmias.
CAUTION
ODO mode is not recommended for pacemaker-dependent patients
or patients who might be affected by even a short cessation of the
device’s operation.
Output Pulse
Programmed Rate
AP
AP
Programmed Rate
AP
AP