Merlin™ PSA Help Manual
3-7
EGM
EGMs (intracardiac electrograms) show the sensed electrical activity of the heart. The shape and size of
the waveform depend on the available
The configurations available are Unfiltered or Sense Amp (narrow band filtered data). The Sense Amp
filtering characteristics of the Merlin PSA are designed to match those of St. Jude Medical™ pulse
generators. The default configuration is Unfiltered, and once it’s changed, the new configuration is
saved for future sessions.
Select the
button to select the waveform source and configuration, and also the sweep
speed and the ECG Filter.
Current of Injury
The Current of Injury view shows an unfiltered view of the EGM and an amplitude value measured
from peak-to-peak of the waveform. (The filtered EGM view measures the Sense Amp waveform from
baseline-to-peak.) The peak-to-peak amplitude of the unfiltered EGM is shown in the upper left corner
of the Current of Injury view. The Current of Injury view updates with every sensed beat.
Use the Plus (+) and Minus (-) buttons to manually set the gain. AutoGain is not available in the
Current of Injury view.
Select the Current of Injury button to return to the Rhythm Display.
Accessed From:
> Channel Control button
Waveform Control
The Waveform Control buttons on the left side of the Rhythm Display control the waveform’s
appearance. To open the controls, select one of the buttons. The buttons shown depend on the
selected Source for that channel.
Figure 3-2. Waveform control buttons
1
button (EGM only). Toggles from the filtered baseline-to-peak view of the
EGM (pictured) to the unfiltered, peak-to-peak Current of Injury view of the EGM.
2
Configuration
button. Opens the
window.
3
AutoGain
button. Allows the programmer to automatically set an appropriate gain.
4
Plus (+)
and
Minus
(-)
buttons allow you to set the gain manually.
Accessed From: Rhythm Display > Channel Control button