8.1 General information
The RIM10 system self-checks its internal status every 500 ms (1/2 second). If there is a
problem with any part of the system, the yellow fault light will appear on the flowmeter front
panel and the main web page screen will show a fault message.
8.2 Web interface Error Checking and Calibration
All faults may be viewed on the Fault Event log through the web user interface.
1.
From the main menu, click the Service button to go to the Service Console page. If a
password has been set, it will need to be entered here (6-digit number).
2.
Click the Inputs or Outputs buttons on the right side of the screen to go to the
Diagnostics / Calibration pages for the input or output data.
3.
Before running any reports or calibrating the inputs or output, you must take the
interface off line. Click the Go Offline button at the top of the Diagnostic, Calibration
page. The fields that were grayed out will now be editable. The green state of the LED
on the flowmeter front panel will stop blinking. You can now calibrate or view reports.
8. Diagnostics and Calibration
Fig. 89
Fig. 90