5.25 Service (calibrate 4-20 mA input / output)
This menu (Figure 56) lets the user calibrate the three 4-20 mA input and output channels.
It requires the user to use an external ammeter to measure the calibration current. See the
Section on calibrating Analog Input and Output - page 84.
Main Menu
Return to Display
Reset Totalisers
Reset Min / Max
Display Set-up
Set-up Wizard
Network Set-up
Set Time / Date
Faults
Service
Service Menu
Reset Password
Calibrate 4-20
Outputs
Factory
Enter / right
arrow
Select channel
Cal ch 1 output
Cal ch 2 output
Cal ch 3 output
Cal ch 1 input
Cal ch 2 intput
Cal ch 3 input
Enter
Terminals :
7 to 13
8 to ammeter -
1 to a
Enter
Output
now set to
approx 3.47 mA
Enter / right
arrow
Enter measured current
3.564
EDIT
Output
now set to
approx 21.50 mA
Enter measured current
21.03
EDIT
Enter
Enter
Connect jump wires between
terminals and ammeter
At this point, the flowmeter
will set the output channel
to 3.47 mA.
Press the enter key
when the Edit field is
selected to change the
measured current in
mA.
Press the enter key when the Edit
field is selected to change the
measured current in mA.
At this point, the flowmeter
will set the output channel
to 21.50 mA.
Fig. 56
Enter
Enter