118153-001 Rev. B
5
1 Scope
This document applies to the communications interface on the uX (Micro-X)
Interface Control Board , assembly 460291-XXX.
2.0 Functional Description
The communications interface provides:
RS-232.
Ethernet
(10/100-Base-T)
USB
Data acquisition and control capabilities are provided by:
6 channels of 12-bit analog-to-digital converters.
2 additional analog channels that monitor the house-keeping 24V power
supply and ambient board temperature
4 channels of 12-bit digital-to-analog converters.
1 digital output bit
3 digital inputs bits
1
relay/interlock
3.0 Getting Started – Hardware Setup
The digital hardware includes a 120MIPS micro control unit processor (MCU).
The uX can operate in manual (analog) control mode or remote (digital control
via the selected interface). The selection is done using DIP Switches located on
the control board.
3.1 DIP
Switches
The DIP switch (SW1) settings determine if the unit is running in local
mode
(analog
control)
or
remote mode (digital control via serial / Ethernet
/
USB).
Switch Settings for Local or Remote Mode
ID
Local Mode
Description
Switch Position
SW1-1
Local
Disable processor KV setting control.
OFF
SW1-2
Local
Bypass processor KV enable relay
ON
SW1-3
Local
Disable processor mA control
OFF
SW1-4
Local
Disable processor filament preheat control.
OFF
SW1-5
Local
Disable processor filament current limit control.
OFF
SW1-6
Local
Enable filament preheat adjustment pot R156.
ON
SW1-7 Local
Enable
filament
current limit adjustment pot
R152.
ON
SW1-8
Local
Spare Not Used.
ON
ID
Remote Mode
Description
Switch Position
SW1-1
Remote
Enable processor KV setting control.
ON
SW1-2
Remote
Enable processor KV enable relay
OFF
SW1-3
Remote
Enable processor mA control
ON
SW1-4
Remote
Enable processor filament preheat control.
ON
SW1-5
Remote
Enable processor filament current limit control.
ON
SW1-6
Remote
Disable filament preheat adjustment pot R156.
OFF
Содержание uX SERIES
Страница 11: ...uX MANUAL 4 118151 001 REV A Figure 2 1 50kV uX DIMENSIONS ...
Страница 12: ...uX MANUAL 5 118151 001 REV A Figure 2 2 65kVuX DIMENSIONS ...
Страница 16: ...uX MANUAL 9 118151 001 REV A Table 3 2 DC Input Filament Connections ...
Страница 17: ...uX MANUAL 10 118151 001 REV A Figure 3 3 Monitors Voltage Current Control Connections ...
Страница 18: ...uX MANUAL 11 118151 001 REV A Figure 3 4 Recommended Interlock HV on Configurations ...