118153-001 Rev. B
38
uX_Manual.doc REV -
6.10 Request Analog Monitor Readbacks
Description:
The host requests that the firmware transmit the present values of the
system analog monitor inputs. Bit scaling in model dependent versions,
see the tables in section 7.0.
Direction:
Host to supply
Syntax:
<STX><20><,><CSUM><ETX>
Example:
<STX>20,<CSUM><ETX>
Response:
<STX><20><,><ARG1><,><ARG2><,><ARG3><,><ARG4><,>
<ARG5><,><ARG6><,><ARG7>,<CSUM><ETX>
Where:
<ARG1> = Control Board Temperature Sensor Reading, range 0-4095
<ARG2> = Low Voltage Supply Monitor, range 0-4095
<ARG3> = kV Feedback Reading, range 0-4095
<ARG4> = mA Feedback Reading, range 0-4095
<ARG5> = Filament Current Reading, range 0-4095
<ARG6> = Filament Voltage Reading, range 0-4095
<ARG7> = High Voltage Board Temperature Sensor, range 0-4095
Example:
<STX>19,500,2048,4095,4095,4095,4095,4095,4095,650,
<CSUM><ETX>