118153-001 Rev. B
41
uX_Manual.doc REV -
6.13
Request Expanded Status
Description:
The host requests that the firmware send and expanded version of the
power supply status.
Direction:
Host to supply
Syntax:
<STX><32><,><CSUM><ETX>
Example:
<STX>32,<CSUM><ETX>
Response:
<STX><22><,><ARG1><,><ARG2><,><ARG3><,><ARG4><,><ARG5><,
><CSUM><ETX>
Where:
<ARG1> 1 = Hv On, 0 = Hv Off
<ARG2> 1 = Interlock 1 Open, 0 = Interlock 1 Closed
<ARG3> 1 = Interlock Fault, 0 = No Interlock Fault
<ARG4> 1 = Over-voltage Fault, 0 = No Over-voltage Fault
<ARG5> 1 = Configuration Fault, 0 = No Configuration Fault
Example Response for HV on, interlock closed, no faults:
<STX>22,1,0,0,0,0,<CSUM><ETX>
Note:
For this command the Interlock, and Over-voltage fault indicators
persist until they are cleared. The Interlock fault (interlock opened while
HV was on) is automatically cleared when the interlock is closed. The
Over-Voltage fault (High Voltage > 106% of unit maximum) is
automatically cleared when High Voltage is again turned on. In addition
the “Clear Faults” command can be sent to manually clear both of these
faults.
A configuration fault indicates an error during power up
configuration of the unit. This can be caused by non-volatile memory
corruption or a hardware failure This fault cannot be cleared using the
reset faults command, and the unit will not allow high voltage to be turned
on until the problem is rectified. Contact Spellman High Voltage.
Содержание uX SERIES
Страница 11: ...uX MANUAL 4 118151 001 REV A Figure 2 1 50kV uX DIMENSIONS ...
Страница 12: ...uX MANUAL 5 118151 001 REV A Figure 2 2 65kVuX DIMENSIONS ...
Страница 16: ...uX MANUAL 9 118151 001 REV A Table 3 2 DC Input Filament Connections ...
Страница 17: ...uX MANUAL 10 118151 001 REV A Figure 3 3 Monitors Voltage Current Control Connections ...
Страница 18: ...uX MANUAL 11 118151 001 REV A Figure 3 4 Recommended Interlock HV on Configurations ...