118153-001 Rev. B
36
uX_Manual.doc REV -
6.8
Request Filament Preheat
Description:
The host requests that the firmware report the current filament preheat
setpoint. The A/bit scaling for the argument is model dependent. See the
tables in section 8 of this document.
Direction:
Host to supply
Syntax:
<STX><16><,><CSUM><ETX>
Response:
<STX><16><,><ARG><,><CSUM><ETX>
Where:
<ARG> = 0 - 4095 in ASCII format
Example Response
<STX>16,4095,<CSUM><ETX>
Содержание uX SERIES
Страница 11: ...uX MANUAL 4 118151 001 REV A Figure 2 1 50kV uX DIMENSIONS ...
Страница 12: ...uX MANUAL 5 118151 001 REV A Figure 2 2 65kVuX DIMENSIONS ...
Страница 16: ...uX MANUAL 9 118151 001 REV A Table 3 2 DC Input Filament Connections ...
Страница 17: ...uX MANUAL 10 118151 001 REV A Figure 3 3 Monitors Voltage Current Control Connections ...
Страница 18: ...uX MANUAL 11 118151 001 REV A Figure 3 4 Recommended Interlock HV on Configurations ...