entered using the standard nomenclature of 4 decimal numeric values separated by
periods, for example “192.168.1.5”.
If the Ethernet settings are lost, forgotten or an error is made when entering, and
you lose access to the uX Ethernet connection, you may restore the above default
settings using the “Reset IP Address” switch on the uX unit. To do this, power down
the uX unit. Push the “Reset IP Address” switch and hold it down, and then power
up the unit while continuing to hold down the switch. Wait 5 seconds while holding
the switch down. Release the switch, power cycle the unit. You should now be able
to access the unit at the default 192.168.1.4 address.
The applet also has an “About” page. This page contains some revision information
related to uX firmware. The information on this page is read only.