118153-001 Rev. B
23
uX_Manual.doc REV -
5.2.1 Ethernet Command Format
<STX><CMD><,>ARG><,><ETX>
Where:
<STX>
= 1 ASCII 0x02 Start of Text character
<CMD>
= 2 ASCII characters representing the command ID
<,>
= 1 ASCII 0x2C character
<ARG>
= Command Argument
<,>
= 1 ASCII 0x2C character
<ETX>
= 1 ASCII 0x03 End of Text character
This is similar to the RS232 serial commands as defined in section 5.1.1, except
the checksum is omitted. The Ethernet interface acts as a transport for the serial
uX commands. The command formats, responses, and examples as shown in
section 5.4 apply to both RS232 and Ethernet communication, except that when
using Ethernet communication, there is no checksum byte.
5.2.2 Ethernet Default IP Address
The default factory settings for the Ethernet interface are:
IP Address – 192.168.1.4
Port Number - 50001
5.2.3 Ethernet Web Server
The uX has resident an applet that can be accessed using a Web Browser such
as Internet Explorer. This can be used to change the Ethernet settings.
To access this applet, open your web browser and in the address bar type
“http://” followed by the IP address, for the factory default settings this would be
“http://192.168.1.4”. The log-on web page should appear in your browser. If you
can’t connect to the uX, verify the cable is attached, the unit is powered and your
network card is set properly (it must be on the same subnet as the uX).
Содержание uX SERIES
Страница 11: ...uX MANUAL 4 118151 001 REV A Figure 2 1 50kV uX DIMENSIONS ...
Страница 12: ...uX MANUAL 5 118151 001 REV A Figure 2 2 65kVuX DIMENSIONS ...
Страница 16: ...uX MANUAL 9 118151 001 REV A Table 3 2 DC Input Filament Connections ...
Страница 17: ...uX MANUAL 10 118151 001 REV A Figure 3 3 Monitors Voltage Current Control Connections ...
Страница 18: ...uX MANUAL 11 118151 001 REV A Figure 3 4 Recommended Interlock HV on Configurations ...