118153-001 Rev. B
39
uX_Manual.doc REV -
6.11
Request Total Hours High Voltage On
Description:
The host requests that the firmware sends the present value of the Total
Hours High Voltage On.
Direction:
Host to supply
Syntax:
<STX><21><,><CSUM><ETX>
Example:
<STX>21,<CSUM><ETX>
Response:
<STX><21><,><ARG1><,><CSUM><ETX>
Where:
<ARG1> is the “HV On” time as a decimal string “X.Y” and
X = hours on.
. = The ASCII Period Character
Y = Tenths of hours.
Example Response: for 1234.9 hours
<STX>21,1234.9,<CSUM><ETX>