Section 5: Options and Accessories
Page 5-2
8197 Instruction Manual
5.1
OPTION 02 INTERNAL BATTERY BACKUP
Option 02 provides the Model 8197 with an internal battery backup power source.
The battery powers the Rubidium oscillator and GPS receiver whenever AC
power fails or is turned off. Battery backup shortens the recovery period to two
hours when AC power returns by eliminating oscillator warm up and GPS
reacquisition. Without backup power, a 3-4 hour recovery period is required.
The battery pack typically provides four hours of backup operation.
NOTE:
The battery backup powers only the oscillator and
GPS receiver during AC power failures. Outputs are not
provided under backup operation.
The battery is continuously float-charged whenever AC power is present.
Maximum recharge period is 20 hours. The battery is tested daily to verify
battery operation. At midnight “receiver time” the microprocessor enables the
battery pack to power the oscillator and GPS receiver for approximately four
minutes. The battery test is transparent to unit operation. If the battery falls
below the minimum voltage set point, the test is discontinued and the Replace
Battery alarm is asserted. Refer to Section 6, Service Information, for battery
replacement instructions.
A battery disconnect switch is provided on the rear panel to prevent battery
discharge when shipping or storing the Model 8197. To disconnect the battery,
first remove the unit from AC power and then depress the recessed switch
labelled BATT DISC.
NOTE:
Failing to disable a battery prior to shipment or
storage may result in reduced capacity and shortened
battery life.
5.1.1 Battery Lamps
Units equipped with Option 02 include front panel indicators to communicate
battery status.
Ready Lamp
: This green lamp is on when the battery is OK and fully charged.
This lamp is off while the battery is tested.
Charging Lamp
: This yellow lamp is on when the battery is charging. This lamp
is off while the battery is tested.
Replace Lamp
: This red lamp turns on and a Minor Alarm asserted when the
battery has remained in charge mode for two days or has failed the daily test.
The battery must be replaced, refer to the service section of this manual.