Section 6: Service Information
Page 6-2
8197 Instruction Manual
7.
Install the replacement battery into the holder and reinstall the
holder cover.
8.
Reconnect the two-position plug to the front panel circuit board.
Verify the plug is properly mated to the header to maintain the
correct polarity.
9.
Reinstall the top cover.
10.
Reapply power to the unit.
NOTE
: The Replace Battery lamp will remain on after
replacing the battery. The lamp can be reset using the
RS-232 command CA BAT, or will self-clear upon
successfully completing the first battery test (typically 48
hours).
To reset the Replace Battery Alarm lamp, a terminal or PC running terminal
emulation software (Hyperterminal, Procomm Plus, etc.) will be needed to
communicate with the RS-232 Comm port. Configure the terminal for ANSI
emulation, 9600 baud, and a character structure of 1 start, 8 data, 1 stop and no
parity. Flow control is not required, though xon/xoff is supported.
The Reset Battery Alarm command is a Set Mode command. Place the unit in
Set Mode by sending the following command:
Type:
SM ON <ent>
The unit will respond with an acknowledgement that Set Mode is enabled. Issue
Battery Reset command as follows:
Type:
CA BAT <ent>
The unit will respond with a status change message to indicate the battery alarm
has been cleared.
This completes the battery replacement procedure. The unit will require a two to
three hour period to obtain oscillator lock to the GPS reference. The Battery
Status lamp will change from Charging to Ready when the battery enters trickle
charge mode.