ST2515 Operation Manual
41
Temperature range display
-99.9°C to 999.9°C
Resolution
1mV
Accuracy
±1%Rd ±3mV
Accuracy= 1% × (T
R
– T
0V
) + 0.3% × (T
1V
- T
0V
)
T
1V
: The temperature measured under input voltage of 1V.
T
0V
: The temperature measured under input voltage of 0V.
T
R
: The current measured temperature.
5.3.5
Temperature correction factor
When the instrument is used under the following environmental conditions:
0°C-18°C, ≤80%RH; 28°C-40°C, ≤80%RH
Measurement Accuracy: Multiply the basic measurement accuracy value listed in chapter 5.3.1 to
5.3.4 by the temperature correction factor K in the following table:
Temperature(°C)
0 - 5
5 - 18
18 - 28
28 - 35
35 - 40
Temperature
correction factor K
4
2
1
2
4
Note:
Rd is the digit reading value of the actual measurement; Fs is the full-scale value of the
selected range.