ST2515 Operation Manual
15
3.4
STATIS DISP
Touch the main menu or press [MEAS]. And then touch the soft key STATIS DISP to enter into the
STATIS DISP page shown as figure 3-4.
Figure 3-4 Statistic display
The function of this page is to count measurement data. It is workable to analyze the average
value of results for multiple measurements, the PASS/FAIL rate and some engineering coefficients.
Details are as below.
3.4.1
Edge Mode and Settings of Its Relevant Values
Touch this key, two modes is available: ABS (high and low limits) and % (percent error mode).
When ABS is selected, the adjacent two items on the same line will be high and low limits. Touch
them to set the values. When % is selected, the adjacent two items on the same line will be
nominal value and percent. Touch them to set the values.
3.4.2
Status ON/OFF
Status ON/OFF:
(1)
When ON is selected, except the trigger button, no other keys and buttons could be enabled.
The instrument makes a statistic at every trigger.
(2)
When OFF is selected, the statistic function will be turned off and other keys and buttons is
available.