ST2515 Operation Manual
13
DATA STORAGE: Save all data.
3.2.3
COMP (ON/OFF)
COMP (ON/OFF): ON indicates the compare function has been turned on. Only when the
compare function is in ON mode, count function can be enabled.
OFF is to turn off this function.
3.2.4
Compare Mode and SETTINGS of ABS & Percent Error
You can set the compare mode and concerned values by touching the screen or pressing
keyboard. There are two compare modes: ABS and %.
Compare modes:
1)
ABS (high and low limits)
In this mode, you can set low and high limits. The instrument will take comparison between the
tested values and the set ones so as to judge the DUT is HI or LO or IN.
2)
% (percent error)
In this mode, you can set the nominal value and the percent error. If you set the nominal value as
100 and percent as 10, the set value should be 100±10%. The instrument will take comparison
between the tested values and the set ones so as to judge the DUT is HI or LO or IN.
3.3
<BIN DISP>
Touch <MEAS DISP> or press [MEAS]. Then select the soft key BIN DISP to enter into <BIN DISP>
shown as figure 3-3.
Figure 3-3 Bin display
The following parameters can be set on <BIN DISP>.
FILE
1)
File Manage: Touch this key to enter into the file manage page.
2)
Screen Copy: Touch this key to save the current screen page to an external U disk. If no U
disk is found, this function will be disabled.