OPERATING MANUAL PAT-800/805 version 1.09
7
Measurement of touch leakage current:
adjustable measurement time
adjustable upper limit in the range of: 0.01 ... 1.99 mA with resolution of 0.01mA
Power measurement:
adjustable measurement time
Current consumption measurement
IEC lead test
Other:
automatic selection of measuring range
990 memory cells for storing individual measurement results with the option to transfer them
to a PC via USB socket or printing
cooperation with the bar-code reader and printer
large, readable display with backlight option
ergonomic operation
Note:
The displayed E02 symbol informs that the 10/25A test set is damaged. The machine
must be sent for repair.
3 Switching on and general settings
3.1 Power supply
The device is powered from the network 187 ... 265V, 50Hz.
Two 15A fuses protect L and N lines from the supply socket to the test socket, they are tripped
when current consumption from the test socket is too high (>16A).
500mA fuse protects 200mA current controller for R
PE
measurement.