OPERATING MANUAL PAT-800/805 version 1.09
51
Measurement of insulation resistance using test voltage of 500V
Test range according to IEC 61557-2 for U
N
= 500V: 500kΩ…599.9MΩ
Display range
Resolution
Measurement uncertainty
(basic)
0...1999kΩ
1kΩ
(5 % m.v. + 8 digits)
2.00 ... 19.99 MΩ
0.01MΩ
20.0...599.9MΩ
0.1MΩ
Influencing factor
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0%
Temperature
E
3
0.1%/ºC
Capacity
E
7
0% for R ≤ 20MΩ
unspecified for R> 20M
Ω
Accuracy of generated voltage (Robc [
]
1000*UN [V]): -0+30% from the set value
nominal current: min 1 ... 1.4 mA
adjustable low
er limit within the range of 0.1 ….9.9 MΩ with resolution of 0,1MΩ
adjustable measuring time: "Cont" mode (continuous measurement) 4s…3min with a
resolution of 1s
detection of a dangerous voltage before commencing a measurement
discharging the object tested
Note:
For R <50k
Ω, the uncertainty is not specified.
Measurement of substitute leakage current
Display range
Resolution
Measurement uncertainty
(basic)
0.00 ... 3.99 mA
0.01 mA
(5% m.v. + 2 digits)
4.0 ... 19.9 mA
0.1 mA
Influencing factor
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0%
Temperature
E
3
0.075%/ºC
opening voltage: 25 ... 50V
internal resistance of the testing device 2k
Ω ± 20%
adjustable upper limit in the range of: 0.01 ... 9.9 mA 0.01 mA with resolution of
0.01mA/0.1mA
adjustable measuring time in the range of: Cont , 1…60s with resolution of 1s
Measurement of PE leakage current
Note
: In the half-time of the measurement, the meter automatically changes the polarity of the test
socket and as a final result it displays the value of higher leakage current.
Display range
Resolution
Measurement uncertainty
(basic)
0.00 ... 3.99 mA
0.01 mA
(5% m.v. + 2 digits)
4.0 ... 19.9 mA
0.1 mA