OPERATING MANUAL PAT-800/805 version 1.09
19
4 Measurements
Notes:
For the convenience of measurements on devices of Class II the socket marked
with
symbol is connected with PE pin of the test socket. Do not connect
dangerous voltage to this socket.
- Tested device must be turned on.
- Measurement activated with defined duration time =
- continuous measurement
– lasts as
long as
START
button is pressed. Maintaining the measurement is possible by pressing
ENTER
with
START
button kept pressed.
- Each measurement with duration =
- continuous measurement
– maintaining the
measurement by pressing
ENTER
may be stopped by pressing
STOP/ESC
.
- After completing each measurement use
,
to see parameters (limits) including the date and
time of measurement.
- All data can be entered using the program on your PC.
4.1 Preliminary test
Connect the mains plug of the tested device into the test
socket of the meter.
Press the button of initial measurement
.
After that meter is prepared for
measurement.