OPERATING MANUAL PAT-800/805 version 1.09
50
Measurement of insulation resistance using test voltage of 100V (only PAT-805)
Test range according to IEC 61557-2 for U
N
= 100V: 100kΩ…99.9MΩ
Display range
Resolution
Measurement uncertainty
(basic)
0...1999kΩ
1kΩ
(5 % m.v. + 8 digits)
2.00 ... 19.99 MΩ
0.01MΩ
20.0 ... 99.9 MΩ
0.1MΩ
Influencing factor
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0%
Temperature
E
3
0.1%/ºC
Capacity
E
7
0% for R ≤ 20MΩ
unspecified for R> 20M
Ω
Accuracy of generated voltage (Robc [
]
1000*UN [V]): -0+30% from the set value
nominal current: min 1 ... 1.4 mA
adjustable lower limit withi
n the range of 0.1 ….9.9 MΩ with resolution of 0,1MΩ
adjustable measuring
time: "Cont" mode (continuous measurement) 4s…3min with a
resolution of 1s
detection of a dangerous voltage before commencing a measurement
discharging the object tested
Note: For R <50k
Ω, the uncertainty is not specified.
Measurement of insulation resistance using test voltage of 250V (only PAT-805)
Test range according to IEC 61557-2 for U
N
= 250V: 250kΩ…199.9MΩ
Display range
Resolution
Measurement uncertainty
(basic)
0...1999k
Ω
1kΩ
(5 % m.v. + 8 digits)
2.00 ... 19.99 MΩ
0.01MΩ
20,0...199,9MΩ
0.1MΩ
Influencing factor
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0%
Temperature
E
3
0.1%/ºC
Capacity
E
7
0% for R ≤ 20MΩ
unspecified for R> 20M
Ω
Accuracy of generated voltage (Robc [
]
1000*UN [V]): -0+30% from the set value
nominal current: min 1 ... 1.4 mA
adjustable lower limit within the range of 0.1 ….9.9 MΩ with resolution of 0,1MΩ
adjustable measuring time: "Cont" mode (continuous measurement
) 4s…3min with a
resolution of 1s
detection of a dangerous voltage before commencing a measurement
discharging the object tested
Note:
For R <50k
Ω, the uncertainty is not specified.