4
cassette could result in unregulated gravity infusion or a
reflux of blood.
• Do not prime the fluid path with the tubing connected to
your catheter. This could result in over-delivery of medica-
tion or air embolism.
• Ensure that the entire fluid path is free of all air bubbles
before connecting to your catheter to prevent air embo-
lism.
• If the pump is dropped or hit, inspect the pump for dam-
age. Do not use a pump that is damaged or is not working
properly. Contact your clinician for further instructions.
Cautions
• Do not operate the pump at temperatures below + 2˚C
(36˚F) or above 40˚C (104˚F).
• Do not store the pump at temperatures below -20˚C (-4˚F)
or above 60˚C (140˚F). Do not store the pump with a
Medication Cassette Reservoir or CADD
™
Administration
Set attached.
• Do not expose the pump to humidity levels below 10% or
above 90% relative humidity.
• Do not store the pump for prolonged periods with the
battery installed.
• Failure to push the Power Pack cord connector or AC
Adapter cord connector all the way forward may result in
an intermittent connection, and the connector may dis-
lodge, causing a loss of power and pump alarms.
• Do not immerse the pump in cleaning fluids or water or